Specifications
System Function:
• Throughput: 600 photometric tests per hour, up to 800 tests per hour with ISE
• Measuring principles: Absorbance Photometry, Turbidimetry
• On-board tests: up to 77 photometric tests + 3 ISEs+ 3 serum indices
• Methodology: End-point, Fixed-time, Kinetic, ISE
Sample Handling:
• SDM: 120 samples by 12 racks, sample continuous loading, STAT in priority
• Sample volume: 1.5~45μL, step at 0.1μL
• HbA1c sample: Whole blood for HbA1c test
• Sample probe: Liquid level detection, clot detection and collision protection
Reagent Handling:
• Reagent tray: 80 positions with cooling systems (2~8°C)
• Reagent volume: 10~200μL, step at 0.5μL
• Reagent probe: Liquid level detection, bubble detection and collision protection
Built-in Bar Code Reader:
• Sample and reagent bar code readers support Codabar, ITF (Interleaved Two of five), Code 128, Code 39, UPC/EAN and Code 93;
• Capable to connect with LIS in the bi-directional mode
Reaction System:
• Reaction volume: 70~300μL
• Reaction temperature: 37°C with 0.1°C fluctuation
• Reaction cuvettes: 124 reusable cuvettes with 8-step auto wash
Optical System:
• Light source: Tungsten-halogen lamp
• Photometer: Grating photometer
• Wavelength: 16 wavelengths (340nm, 380nm, 412nm, 450nm, 480nm, 505nm, 546nm, 570nm, 605nm, 630nm, 660nm, 700nm, 740nm, 770nm, 800nm, 850nm)
• Absorbance range: 0~3.5Abs (10mm conversion)
Control and Calibration:
• Calibration mode: Linear (one-point, two point and multi-point), Logit-Log 4P, Logit-Log 5P, Spline, Exponential, Polynomial, Parabola
• Control rules: Westgard multi-rule, Twin plot
Operation Unit:
• Operation system: Touch screen (optional), Windows 10 (64bit)
• Interface: RS-232, Network Port, USB/ parallel port
Working Conditions:
• Power Supply: 200~240V, 50/60Hz, 1700VA or 110/115V, 60Hz, 1700VA
• Temperature: 15~30°C
• Humidity: 35~85%
• Dimension: 1380mm (length)*860mm (width)*1200mm (high)
• Weight: 350kg